I'm not able to select the advertised sample rate in the software, what's wrong?

Follow

For some Saleae devices, the maximum sample rates are only available in certain configurations:

For Logic Pro 8 and Logic Pro 16, the maximum digital sample rate (500 MSPS) can only selected when only 4 or fewer digital channels are active, and all analog channels are disabled.

If you are still not able to select 500 MSPS, it could be caused by using a USB 2.0 connection instead of a USB 3.0 connection. Review this article: The Saleae Logic Pro 8 or Logic Pro 16 is only operating over USB 2.0 on a USB 3.0 port, or any Saleae device does not work properly on a USB 3.0 port, but does work over USB 2.0

For the new Logic 8, the maximum digital sample rate of 100 MSPS can be selected when 3 or fewer digital channels, and no analog channels.

For Logic 4, there are no restrictions.

For the original Logic16, the maximum sample rate of 100 MSPS can only be selected when 3 or fewer channels are active.

The original Logic has no restrictions.

If you can select a sample rate, but are unable to record at that sample rate (the capture ends with an error) please see this article: The device is not successfully capturing data consistently